Hanna Instruments Taiwan
The Hanna Advantage: A Worldwide Leader in Technology and Innovation 100% Quality  360 Value
Contact Hanna: +886-2-8797-2918
Hanna Instruments Worldwide
Customer Support: +1 (800) 426-6287

Multiparameter pH/mV/ISE and EC/TDS/Salinity Benchtop Meter - HI2550

The HI2550 is a rugged and highly accurate multiparameter benchtop meter. Offering two input channels, this meter is able to measure up to seven parameters including pH, ORP, ISE, EC, TDS, salinity, and temperature. The meter comes with Hanna’s HI1131B glass pH electrode, HI76310 four-ring conductivity probe, and a temperature probe to allow for automatic temperature compensation. The HI2550 offers many features including five-point pH calibration, auto ranging EC and TDS measurements, data logging, comprehensive GLP data, and PC connectivity via USB to provide users with a simple way to organize and manage their data.

  • Automatic Temperature Compensation
  • Auto Ranging EC and TDS measurements
  • Two Independent Channels

Description

The HI2550 is a rugged and highly accurate multiparameter benchtop meter that measures pH/ORP/ISE, EC/TDS/Salinity, and temperature with a clear LCD display, data logging capability, and USB port for computer connectivity.

The HI2550 is a dual channel meter that allows for measurement of pH, ORP, and ISE on one channel and EC, TDS, and salinity, on the other. Readings for pH and EC can be manually or automatically compensated for temperature variations with the separate HI7662 temperature probe or from the built-in temperature sensor of the conductivity probe on channel two.

As a pH meter, the HI2550 can be calibrated up to five points with a choice of seven standard buffers or two custom buffers. As an EC/TDS/Salinity meter, the HI2550 can be calibrated at one point for offset and at one point for slope with a choice of six conductivity standards. The % NaCl is calibrated to a single point at 100% with the optional HI7037 salinity standard. The calibration data including date, time, standards used, offset, slope, cell constant, and cell factor can be accessed at any time by pressing the dedicated GLP button.

Features at-a-glance

HI1131B pH Electrode – The HI2550 is supplied with the HI1131B glass body, double junction, refillable pH electrode with an indicating sensor made of High Temperature (HT) glass. The double junction and HT glass design allows the HI1131B to be used in a wide variety of application from samples with metals and Tris buffer to samples at elevated temperatures.

HI76310 Conductivity Probe – The HI76310 platinum, four-ring probe measures EC, TDS, and salinity. The probe features a built-in temperature sensor and operates over a wide range from 0.00 μS/cm to 500.0 mS/cm (actual conductivity), 0.00 to 400.0 g/L (actual TDS), and 0.0 to 400.0% NaCl.

Temperature Compensation – Temperature measurements for pH and EC can be compensated for automatically (ATC) or manually (MTC) from -20.0 to 120.0℃, or compensation can be disabled for actual conductivity or TDS measurements. The temperature correction coefficient for EC or TDS, also referred to as β, is adjustable from 0.00 to 6.00 %/℃.

Automatic Calibration – The pH channel offers up to five-point pH calibration with seven standard buffers and up to two custom buffers. EC and TDS are automatically calibrated at one point with a choice of six pre-programmed standards. Salinity is calibrated at one point using the HI7037 100% NaCl standard solution.

Autoranging – In the EC and TDS measurement modes, the HI2550 automatically chooses the optimum measurement scale based on the sample solution to maintain the highest possible accuracy.

Adjustable TDS Factor – The factor that relates conductivity to total dissolved solids is based on the type of sample being measured. For users to get an accurate determination of TDS based on their unique solution, the TDS factor is adjustable from 0.40 to 0.80.

GLP Data – The calibration data including date, time, standards used, offset, slope, cell constant, and cell factor can be accessed at any time by pressing the dedicated GLP button.

Data Logging – The log-on-demand feature allows up to 200 data points to be recorded. Up to 500 data points can be recorded when utilizing interval logging, which allows the user to specify time intervals from 5 seconds to 180 minutes.

Data Transfer – With a built-in logging function, measurements are stored in non-volatile memory, and can be transferred to a PC through the USB port using the optional HI92000 software and HI920013 USB cable (both sold separately).

Additional information

Weight 3.05 kg
Parameter

Conductivity/TDS, ORP, pH, Salinity

pH Accuracy

±0.002 pH, ±0.01 pH

pH Calibration

10.01, 12.45), 4.01, 6.86, 7.01, 9.18, and two custom buffers, seven standard buffers available (1.68, up to five point calibration

pH Range

-2.00 to 16.00 pH, -2.000 to 16.000 pH, –2.0 to 16.0 pH

pH Resolution

0.001 pH, 0.01, 0.1

EC Accuracy

±1% of reading (±0.05 µS/cm or 1 digit, whichever is greater)

EC Calibration

111.8 mS/cm; one point offset: 0.00 µS/cm, 12.88, 1413 µS/cm; 5.00, 80.0, one point slope calibration; six buffers available: 84.0

EC Range

0.00 to 29.99 µS/cm, 3.00 to 29.99 mS/cm, 30.0 to 200.0 mS/cm, 30.0 to 299.9 µS/cm, 300 to 2999 µS/cm, up to 500.0 mS/cm actual* conductivity

EC Resolution

0.01 mS/cm, 0.01 µS/cm, 0.1 mS/cm, 0.1 µS/cm, 1 µS/cm

TDS Accuracy

±1% of reading (±0.03 ppm or 1 digit, whichever is greater)

TDS Range

0.00 to 14.99 ppm (mg/L), 1.50 to 14.99 g/L, 15.0 to 100.0 g/L, 15.0 to 149.9 ppm (mg/L), 150 to 1499 ppm (mg/L), up to 400.0 g/L actual* TDS (with 0.80 factor)

TDS Resolution

0.01 g/L, 0.01 ppm, 0.1 g/L, 0.1 ppm, 1 ppm

Temperature Accuracy

0.1°C (0.1°F)

Temperature Range

-20.0 to 120.0°C (-4.0 to 248.0°F)

Specifications

SKU HI2550-01
Name Multiparameter pH/mV/ISE and EC/TDS/Salinity Benchtop Meter – HI2550
pH Range –2.0 to 16.0 pH, -2.000 to 16.000 pH, -2.00 to 16.00 pH
pH Resolution 0.1, 0.01, 0.001 pH
pH Accuracy ±0.01 pH, ±0.002 pH
pH Calibration up to five point calibration, seven standard buffers available (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45), and two custom buffers
pH Temperature Compensation automatic or manual from -20.0 to 120.0 °C
mV Range ±999.9 mV; ±2000 mV (ISE & ORP)
mV Resolution 0.1 mV (±1000.0 mV); 1 mV (± 2000 mV)
mV Accuracy ± 0.2 mV (±999.9 mV); ± 1 mV (±2000 mV)
mV Relative Offset ±2000 mV
EC Range up to 500.0 mS/cm actual* conductivity, 0.00 to 29.99 µS/cm, 30.0 to 299.9 µS/cm, 300 to 2999 µS/cm, 3.00 to 29.99 mS/cm, 30.0 to 200.0 mS/cm
EC Resolution 0.01 µS/cm, 0.1 µS/cm, 1 µS/cm, 0.01 mS/cm, 0.1 mS/cm
EC Accuracy ±1% of reading (±0.05 µS/cm or 1 digit, whichever is greater)
EC Calibration one point slope calibration; six buffers available: 84.0, 1413 µS/cm; 5.00, 12.88, 80.0, 111.8 mS/cm; one point offset: 0.00 µS/cm
TDS Range up to 400.0 g/L actual* TDS (with 0.80 factor), 0.00 to 14.99 ppm (mg/L), 15.0 to 149.9 ppm (mg/L), 150 to 1499 ppm (mg/L), 1.50 to 14.99 g/L, 15.0 to 100.0 g/L
TDS Resolution 0.01 ppm, 0.1 ppm, 1 ppm, 0.01 g/L, 0.1 g/L
TDS Accuracy ±1% of reading (±0.03 ppm or 1 digit, whichever is greater)
EC/TDS Temperature Compensation automatic or manual from -20.0 to 120.0°C, or disabled
Temperature Correction Coefficient 0.00 to 6.00 %/°C (for EC and TDS only; default value is 1.90 %/°C
EC to TDS Conversion Factor 0.40 to 0.80 (default value is 0.50)
Salinity (% NaCl) Range 0.0 to 400.0% NaCl
Salinity (% NaCl) Resolution 0.1% NaCl
Salinity (% NaCl) Accuracy ±1% of reading (excluding probe error)
Salinity (% NaCl) Calibration one point, with HI7037 calibration solution (optional)
Temperature Range -20.0 to 120.0°C (-4.0 to 248.0°F)
Temperature Resolution ±0.4°C (±0.8°F) (excluding probe error)
Temperature Accuracy 0.1°C (0.1°F)
pH Electrode HI1131B glass body pH electrode with BNC connector and 1 m (3.3’) cable (included)
EC/TDS Probe HI76310 platinum four-ring EC/TDS probe and 1 m (3.3’) cable (included)
Temperature Probe HI7662 temperature probe, stainless steel with 1 m (3.3’) cable (included)
Logging Type log-on-demand;
Logging Memory 200 samples
Logging Interval 500 records; 5, 10, 30 sec and 1, 2, 5, 10, 15, 30, 60, 120, 180 min stability logging
Connectivity opto-isolated USB
Power Supply 12 VDC adapter (included)
Environment 0 to 50°C (32 to 122°F), RH max 95% non-condensing
Dimensions 235 x 222 x 109 mm (9.2 x 8.7 x 4.3”)
Weight 1.3 kg (2.9 lb); kit with holder 2.1 kg (4.6 lb)

Ordering Info

HI2550 is supplied with HI1131B pH electrode, HI76310 EC/TDS probe, HI7662 temperature probe, HI76404N electrode holder, HI70004 pH 4.01 buffer solution sachet, HI70007 pH 7.01 buffer solution sachet, HI7071 Electrolyte Fill Solution, 3.5M KCl + AgCl (30 mL), 12 VDC adapter and instruction manual.

Manuals

Multiparameter pH/mV/ISE and EC/TDS/Salinity Benchtop Meter – HI2550