Mini Titrator for Measuring Titratable Acidity in Water - HI84530U

The HI84530 is an easy to use, fast and affordable, method specific titrator designed for testing titratable acidity levels in water. Based on an acid-base titration method, this mini titrator uses an optimized pre-programmed method of analysis with a powerful algorithm that determines the completion of the titration reaction potentiometrically with a glass body pH electrode.

The HI84530 incorporates a precision piston style dosing pump that adjusts the volume of dosing dynamically based on the voltage change. This type of dosing system reduces the amount of time required for the titration while providing for a highly accurate determination of the amount of titrant used.

This mini titrator is supplied complete with all the materials necessary to perform low and high range measurements of acidity in water. All chemicals are premixed and prepackaged including standardized titrants, reagents and pump calibration solution. There is no need for volumetric glassware or analytical balances.

  • Piston driven pump for accurate dynamic dosing
  • Pre-standardized titrant and pre-measured reagents
  • Benchtop use as a pH/mV meter

Description

Acidity is an important parameter used to determine the quality of water (surface, drinking, wastewater). Acidity affects many things including chemical reaction rates, biological processes, and corrosiveness. It can also be used to monitor pollution in wastewater and drinking water.

Total titratable acidity is a quantitative measurement of the water sample’s ability to react with a strong base solution to a specific pH value. Many chemical species can contribute to the acidity measurement depending on the method of analysis. These species can include strong acids (hydrochloric acid, nitric acid, sulfuric acid, etc.), weak acids (acetic acid, carbonic acid, etc.), and hydrolyzing salts (iron, aluminum, manganese, etc.)

Double Junction pH Electrode

The HI84530 is supplied with the HI1131B refillable, double junction, combination pH electrode. By design, the HI1131B has a spherical tip for use in aqueous or liquid solutions. This versatile electrode provides a wide surface of contact with a sample and is ideal for direct measurements or acid base titrations in the water industry.

Piston Driven Dosing Pump

The heart of the HI84530 is the piston driven burette pump. This type of dosing system uses a motor in which each dose is very precisely controlled and the volume dispensed is accurately determined. The piston driven burette is controlled dynamically so that the volume of titrant being dosed is automatically adjusted based on a voltage response of the previous dose. This type of dosing speeds up the titration process by allowing for more titrant to be dosed at the start of the titration and then very small doses as the endpoint is reached.

Automatic Stirrer

The built-in stirrer is automatically maintained at a speed of 600 rpm, regardless of the viscosity of the solution being titrated.

On-Screen Features

titrator screenshot

Tutorial and Help ScreensThe HI84530 has an intuitive user design with clearly defined keys and a large screen that is easy to navigate. The meter has a built in tutorial mode that, when enabled, will walk the user step by step through the titration process. A dedicated HELP key is always available to allow access to content-specific information during both calibration and titration.

Titrate LR

Procedure Warnings Users are warned if there is an error in procedures such as the titration exceeded the maximum volume of titrant.

mini titrator acidity graphic mode

Titration Curve Displayed on Screen This mini titrator displays in-depth data during titration, including a real-time graph of the titration curve.

mini titrator acidity logging

Log and Recall Data The HI84530 allows for data logging of up to 400 samples: 200 titration results and 200 pH/mV readings. Data can be stored and exported to a USB drive or a PC using the USB connection.

mini titrator pH mv meter

pH/mV Meter In addition to being an automatic titrator, the HI84530 also can be used as a pH/mV meter. As a pH meter, it has many features of a professional grade benchtop including automatic calibration up to three points with four available buffers, a 0.01 pH resolution, accuracy of ±0.01 pH, automatic temperature compensation and comprehensive GLP data.

mini titrator acidity graphic mode

CAL Check™ Accuracy is always ensured with Hanna’s unique CAL Check feature, which analyzes the response of the electrode during the calibration process. Based on electrode response in the buffer, indicators are displayed on screen to alert the user of potential problems during calibration. These indicators include Buffer Contaminated, Electrode Dirty/Broken, and overall probe condition. The CAL Check function not only ensures an accurate pH reading when the HI84530 is used as a pH meter, but also an accurate titration since the endpoint of a titratible acidity titration is determined by a set pH value.

Additional information

Weight 5.4 kg
Parameter

Acidity, pH, Titration

pH Accuracy

±0.01 pH (@25ºC/77ºF)

pH Calibration

10.01), 7.01, 8.30, one, two or three-point calibration; four available buffers (4.01

pH Range

-2.0 to 16.0 pH; -2.00 to 16.00 pH

pH Resolution

0.1 pH / 0.01 pH

Temperature Accuracy

±0.4°C; ±0.8°F; ±0.4 K (@25ºC/77ºF), without probe error

Temperature Range

-20.0 to 120.0°C; -4.0 to 248.0°F; 253.2 to 393.2 K

Specifications

SKU HI84530U-01
Name Mini Titrator for Measuring Titratable Acidity in Water – HI84530U
pH-mV Range -2000.0 to 2000.0 mV
pH-mV Resolution 0.1 mV
pH-mV Accuracy ±1.0 mV (@25ºC/77ºF)
pH Range -2.0 to 16.0 pH; -2.00 to 16.00 pH
pH Resolution 0.1 pH / 0.01 pH
pH Accuracy ±0.01 pH (@25ºC/77ºF)
pH Calibration one, two or three-point calibration; four available buffers (4.01, 7.01, 8.30, 10.01)
pH Temperature Compensation manual or automatic
Temperature Range -20.0 to 120.0°C; -4.0 to 248.0°F; 253.2 to 393.2 K
Temperature Resolution 0.1°C; 0.1°F; 0.1K
Temperature Accuracy ±0.4°C; ±0.8°F; ±0.4 K (@25ºC/77ºF), without probe error
Titration Method acid base titration, total acidity / strong acidity
Titration Principle fixed endpoint titration: 8.30 pH (phenolphthalein) or 3.7 pH (methyl orange)
Titration Range Low Range (as CaCO₃): 15.0 to 400.0 mg/L; 0.3 to 8.0 meq/L
High Range (as CaCO₃): 300 to 4000 mg/L; 6.0 to 80.0 meq/L
Titration Resolution Low Range: 0.1 mg/L / 0.1 meq/L; High Range: 1 mg/L / 0.1 meq/L
Titration Accuracy Low Range: ±0.5 mg/L or 3% of reading, whichever is greater (@25ºC/77ºF)
High Range: ±15 mg/L or 3% of reading, whichever is greater (@25ºC/77ºF)
Sample Size Low Range: 50 mL; High Range: 50 mL
Titration Pump Speed 10 mL/min
pH Electrode HI1131B glass body, refillable, with BNC connector and 1 m (3.3’) cable (included)
Temperature Probe HI7662-T stainless steel temperature probe with 1 m (3.3’) cable (included)
Logging Type Log-on-demand
Logging Memory up to 400 ( 200 titrations, 200 pH/mV readings)
Connectivity 1 USB port for data storage and export to a PC
Built-in Magentic Stirrer Built-in magnetic stirrer at 600 rpm
Power Supply 12 VDC adapter (included)
Environment 0 to 50°C (32 to 122°F); RH max 95% non-condensing
Dimensions 235 x 200 x 150 mm (9.2 x 7.9 x 5.9”)
Weight 1.9 kg (67.0 oz.)

Ordering Info

HI84530 is supplied with the HI1131B pH electrode, HI7662-T temperature probe, HI84530-50 low range titrant, HI84530-51 high range titrant, HI84530-55 calibration standard, HI84530-60 hydrogen peroxide, 100 mL beakers (2), dosing pump valve, 5 mL syringe, 1 mL plastic pipette, tube set (aspiration tube with titrant bottle cap and dispensing tube with tip), stir bar, power adapter, instruction manual and quality certificate.

Manuals

Mini Titrator for Measuring Titratable Acidity in Water – HI84530U