Iodide Combination Ion-Selective Electrode (ISE) - HI4111

The HI4111 is a solid-state, combination ion-selective electrode (ISE) for the determination of iodide (I) in solution. The silver iodide sensing pellet produces a potential change due to the iodide ion exchange between the inorganic pellet membrane and the sample. The internal sensing elements are housed within a durable polyetherimide (PEI) body.

The HI4111 is ideal for a variety of applications in food production and environmental monitoring.

  • Solid state sensor
  • Skirted cone geometry
  • Detection from 0.01 to 127, 000 mg/L I

Description

The HI4111 iodide selective electrode is a complete potentiometric cell that contains a silver/silver chloride (Ag/AgCl) reference housed within a PEI plastic body with a potassium nitrate electrolyte. The internal elements are isolated from the sample by a compressed silver iodide sensing membrane.

Hanna’s solid state combination ISEs are designed with a cone style junction. The skirted cone geometry of the sensor forms the liquid junction with the test solution at the tip of the junction cone, producing a highly stable reference potential and consistent flow rates of electrolyte into the solution. A simple press of the top portion of the ISE makes it easy to empty the electrolyte fill solution and replace when necessary.

For the HI4111 to measure iodide accurately, it is important that the ionic strength of the standards and sample be adjusted to a high and constant value with the halide ionic strength adjuster (ISA) solution. The halide ISA ensures that the activity coefficient is constant so the concentration of free iodide ions can be measured. Measuring all solutions with a constant ionic strength reduces the margin of error between measurements. The HI4111 iodide solid state ISE can measure from 0.1µM (0.01 mg/L) to1M (127, 000 mg/L).

Conversion Table for I

Multiply

moles/L (M) to ppm (mg/L)

1.269 x 105

ppm (mg/L) to M (moles/L)

7.88 x 10-6

Features at-a-glance

Fixed Sensing Modules – Combination solid state ISEs are electrodes that detect free ions in a solution. The highly insoluble, compressed sensing pellet is attached to the electrode by a fixed sensing module for ease of use.

Durable PEI Body – The probe body is composed of durable polyetherimide (PEI) plastic.

BNC Connection – The HI4111 has a universal BNC connector for easy connection to any benchtop meter with a female BNC probe input.

Theory of Operation

A solid state sensor electrode develops a voltage due to the ion exchange that occurs between the solution and the inorganic membrane. An equilibrium occurs due to the highly limited solubility of the membrane material in the sample. When the ionic strength of the solution is fixed by the addition of ISA, the voltage is proportional to the concentration of free ions in solution.

Additional information

Weight 0.45 kg
Parameter

Iodide

Temperature Range

0 to 80 °C

Specifications

SKU HI4111
Name Iodide Combination Ion-Selective Electrode (ISE) – HI4111
Description combination
Measurement Range iodide (I) 1M to 1 10-7M, 127,000 to 0.01 mg/L (ppm)
Temperature Range 0 to 80 °C
Body Material PEI
Body Length/Overall Length 120 mm
Outer Diameter 12 mm
Cable coaxial; 1 m (3.3’)
Connector Type BNC
Notes Approximate Slope -56 mV

Manuals

Iodide Combination Ion-Selective Electrode (ISE) – HI4111