Hanna Instruments Taiwan
The Hanna Advantage: A Worldwide Leader in Technology and Innovation 100% Quality  360 Value
Contact Hanna: +886-2-8797-2918
Hanna Instruments Worldwide
Customer Support: +1 (800) 426-6287

Chlorine, pH, ORP, and Temperature Analyzers - PCA300 Series-PCA320 - Chlorine, pH & Temperature

The Hanna Instruments PCA310, PCA320, PCA330 and PCA340 series of chlorine, pH, ORP and temperature analyzers are microprocessor controlled, process analyzers which continuously monitor a sample stream for chlorine content, pH, ORP and temperature values.

  • 4 models available for monitoring chlorine, pH, and ORP levels.
  • Two separate selectable 0-20 or 4-20 mA signal output (PCA340 only)
  • User-friendly interface simplifies setup and maintenance.

Description

PCA chlorine analyzer

Chlorine

From drinking and wastewater treatment to pool and spa sanitation, monitoring chlorine levels has an important bearing on public health as well as ROI and efficiency for heating system and industrial applications.

As one of the oldest and most common forms of disinfection, chlorine improves water quality by destroying disease-producing microorganisms, and by reacting with other organic and inorganic substances. Chlorine levels must be actively monitored to ensure sufficient chlorine is present for disinfection, as well as to control adverse effects such as taste, odor, and potential reactions with organic matter to form harmful disinfection byproducts.

Chlorine is a strong oxidising agent that destroys mostly organic pollutants and bacteria and can combine with nitrogen containing compounds, forming chloramines. When dosing chlorine for disinfection, only a portion of the dosed chlorine remains active to actually continue the disinfection process.

PCA chlorine Analyzer

DPD Colorimetric method

The PCA Analysers uses DPD Colorimetric method in which N, N-Diethyl-p-phenylenediamine indicator and a buffer are mixed together with the sample. The resulting chemical reaction causes a magenta color to form in the presence of chlorine. The color intensity is proportional to the concentration. The color intensity is measured photometrically (light source at a specific wavelength and a photodetector) and converted to chlorine concentration, in mg/L, which is displayed on the front panel. The sampling interval for chlorine measurement is adjustable from 3 to 90 minutes. The PCA has a dosing relay for the addition of chlorine by a dosing pump or chlorine generator when a reading is below the programmable set point. The technology used by the PCA for chlorine measurement is the same as that found in portable and benchtop colorimeters providing for consistent results when performing process verification with one of those types of meters.

Amplified pH electrode

The PCA Family also uses the HI1005 amplified pH electrode with a built in pt100 temperature sensor and matching pin to measure both pH and temperature. The built in amplifier and matching pin provide for exceptional performance against any electrical noise generated by pumps and motors.

Analog output

The PCA340 features two selectable 0-20 or 4-20 mA signal output that are scalable for the transmission of readings to external recording devices. The analog outputs can also be set for dosing and used with dosing pumps that accept a 4-20 mA analog input. The analog outputs can be used for any of the three measured parameters.

PCA Series Controller FEATURES/BENEFITS:

Backlit LCD Display

  • The PCA Family has a backlit display that is easy to read from a distance and allows for up to three parameters to be displayed at a time.

Nema 4X Protection

  • The PCA Analysers are enclosed in waterproof casing for superior protection against the elements. The front door of the case has a window for the measurement display while also shielding the DPD reagents from UV light to prevent premature degradation.

DPD Chlorine Measurement Method

  • The DPD colorimetric method is one of the most common and reliable methods to measure chlorine. The PCA Family can use either free or total chlorine reagents and allow for 16, 000 measurements to be performed.

Colorimeter Diagnostics

  • Advanced diagnostics allow for easy troubleshooting of the colorimeter. In the setup menu it is possible to select an option that allows the user to determine the difference between a dark read (LED off) and a blank read (LED on). These Analuzers also automatically performs this check in order to determine when to alert the user that the sample cell needs to be cleaned.

Reagent Reminder

  • The PCA family has a reagent reminder feature to alert the user when the reagents are running low. When the reagents are changed the counter is reset and the meter automatically tracks the number of readings performed.

Amplified pH/Temperature Probe

  • An integrated pt100 temperature sensor allows for automatic temperature compensation of pH measurements and allows for monitoring temperature as well. The built in amplifier and matching pin provides for exceptional performance where other probes fail when placed in line with pumps and motors.

Data Logging

  • The analyzers can store up to 3500 readings (at least 7 days worth of records when set to a a3 minutes sampling interval) that can be reviewed or downloaded to a Windows compatible PC using the HI92500 software and the RS485 serial port. Logged records contain the date time and reading of all parameters measured along with any alarm status.

GLP Data

  • The GLP data allows for the user to review the data and time for the last Chlorine and pH calibration.

Digital RS485 Output

  • The PCA Family has a RS485 digital output that allows for connection to a Windows compatible PC running the HI92500 software. The software allows for remote monitoring, review of logged data, events and errors, and executing setup options.

Two Analog Outputs (PCA340)

  • The PCA340 features two selectable 0-20 or 4-20 mA signal output that are scalable for the transmission of readings to external recording devices. The analog outputs can also be set for dosing and used with dosing pumps that accept a 4-20 mA analog input. The analog outputs can be used for any of the three measured parameters.

Two Dosing Relays

  • The dosing relays of these analyzers can be connected to a pH and/or chlorine dosing pumps. The chlorine relays are proportionally controlled while the pH relay can be set for on/off or proportional control. The proportional controloffers very fine control of doing to prevent any
    overshoot and wastage of chemicals.

Alarm Relay

  • One SPDT alarm relay is provided that can be activated by adjustable upper and lower chlorine, pH and temperature limits.

Error Relay

  • One SPDT error relay is provided and is activated when an error is present including a problem with the colorimeter such as when the reagent counter has reached zero, or when a reading is outside the range for a measured parameter.

Warning Messages

  • Error messages are displayed when the reagents are expired or low and if the colorimeter cell needs to be cleaned.

Additional information

Weight 8.81 kg
Parameter

Chlorine, ORP, pH

pH Accuracy

(PCA320, PCA330 and PCA340): ±0.05 pH

pH Calibration

automatic one or two points; one point process calibration

pH Range

(PCA320, PCA330 and PCA340): 0.00 to 14.00 pH

pH Resolution

(PCA320, PCA330 and PCA340): 0.01 pH

Temperature Accuracy

(PCA320, PCA330 and PCA340): ±0.5 °C (±1.0 °F)

Temperature Range

(PCA320, PCA330 and PCA340): 5.0 to 75.0 °C (41.0 to 167.0 °F)

Specifications

SKU PCA320-1
Name Chlorine, pH, ORP, and Temperature Analyzers – PCA300 Series-PCA320 – Chlorine, pH & Temperature
pH Range (PCA320, PCA330 and PCA340): 0.00 to 14.00 pH
pH Resolution (PCA320, PCA330 and PCA340): 0.01 pH
pH Accuracy (PCA320, PCA330 and PCA340): ±0.05 pH
pH Calibration automatic one or two points; one point process calibration
pH Dosing Relay ON/OFF or proportional, relay or 4-20mA output
pH Dosing Set Point selectable from 0.10 to 2.00 pH
pH Hysteresis Band selectable from 0.05 to 2.00 pH
pH Sampling Rate adjustable from 3 to 120 seconds
ORP Range (PCA330): 0 to 2000 mV
ORP Resolution (PCA330): 1 mV
ORP Accuracy (PCA330): ±1 mV
Temperature Range (PCA320, PCA330 and PCA340): 5.0 to 75.0 °C (41.0 to 167.0 °F)
Temperature Resolution (PCA320, PCA330 and PCA340): 0.1 °C (0.1 °F)
Temperature Accuracy (PCA320, PCA330 and PCA340): ± 0.5 °C (± 1.0 °F)
Free Chlorine Range 0.00 to 5.00 mg/L (ppm)
Free Chlorine Resolution 0.01 mg/L (ppm)
Free Chlorine Accuracy ± 8% or ±0.05 mg/L whichever is greater
Total Chlorine Range 0.00 to 5.00 mg/L (ppm)
Total Chlorine Resolution 0.01 mg/L (ppm)
Total Chlorine Accuracy ± 8% or ±0.05 mg/L whichever is greater
Chlorine Calibration One point process calibration
Chlorine Sampling Rate adjustable from 3 to 90 minutes
Chlorine Dosing Relay proportional relay or 4-20 mA output
Chlorine Dosing Set Point selectable from 0.1 to 5 mg/L (ppm)
Dosing Control SPDT contact with 5A, 230V resistive load
Alarm Relay Set Point SPDT contact with 5A, 230V resistive load
Analog Ouput Signal 0-10 mV, 0-100 mV, 0-1 V, 4-20 mA, 0-20 mA (PCA310, PCA320, PCA330) 4-20 mA, 0-20 mA (PCA340)
Baud Rate 1200; 2400; 4800; 9600 bps
Enclosure NEMA-4X standard, molded fiberglass polyester with transparent Lexan window
Inlet Sample Pressure 0.07 to 4 bar with no external pressure regulator (for pressure exceeding four bar an external pressure regulator is required)
Sample Flow Rate 100 to 300 mL/min
Sample Temperature 5 to 40°C (41 to 104°F)
Sample Inlet/Outlet Connection 12mm (1/2”) male NPT fitting
Drain Connection 10mm (3/8 in.) barb
Logging Memory 3500 log records
Connectivity RS485, galvanic separated
Display character LCD 4 lines x 20 characters (languages: English, Italian, Portuguese, Spanish)
Power Supply 115 VAC ±10% or 230 VAC ±10%; 50/60 Hz; 20 VA
Dimensions 318 x 267 x 159 mm (12.5 x 10.5 x 6.25”)
Weight 5 kg (11 lb.) without reagents

Ordering Info

Each analyzer is supplied complete with: 2 pcs. reagent bottles (1 pc. indicator and 1 pc. buffer solution), 2 pcs. reagent bottle caps, 5 pcs. DPD compound powder, tubing, and instruction manual.

Manuals

Chlorine, pH, ORP, and Temperature Analyzers – PCA300 Series-PCA320 – Chlorine, pH & Temperature