Hanna Instruments Taiwan
The Hanna Advantage: A Worldwide Leader in Technology and Innovation 100% Quality  360 Value
Contact Hanna: +886-2-8797-2918
Hanna Instruments Worldwide
Customer Support: +1 (800) 426-6287

Marine Master Waterproof Wireless Multiparameter Photometer | HI97115

marine

pH, Alkalinity, Ammonia, Calcium, Magnesium,
Nitrate Low Range, Nitrate High Range, Nitrite
Ultra Low Range, Phosphate Ultra Low Range

The HI97115 is a compact and versatile Marine multiparameter photometer designed to accurately determine pH, Alkalinity, Ammonia, Calcium, Magnesium, Nitrate, Nitrite, and Phosphate levels in aquariums and marine biology applications. The HI97115 is suitable for field and bench measurements.

Description

Marine, Saltwater Aquariums, Aquaculture

Precise and advanced LED optical system

The HI97115 can be used as a stand-alone photometer or can be connected to the Hanna Lab App with a compatible smart device via the integrated Bluetooth module. When connected, Hanna Lab App functions include measurement with the ability to add notes, data logging with extended storage capacity, data sharing, and the ability to create and save method groups.

  • Precise and advanced LED optical system
  • Waterproof and floating IP67 meter design
  • Meter validation using certified CAL Check™ cuvettes
  • Includes auto data logging features to easily record up to 200 water testing results
  • On-screen tutorial mode guides users step-by-step through the measurement process
  • Option to assign locations to logged readings
  • Positive-locking system ensures that the cuvettes are placed into the holder in the same position every time
  • Future compatibility with Hanna Cloud

Innovative optical design that utilizes a reference detector and focusing lens to eliminate errors from changes in the light source and from imperfections in the glass cuvette.

LEDs have a much higher luminous efficiency, providing more light while using less power. They also produce little heat, which could otherwise affect electronic stability.

diag

CAL Check™ Validation

Hanna’s exclusive CAL Check™ feature provides a stress-free, straightforward way to ensure your photometer is working properly. Simply insert the certified CAL Check™ standards and let your meter guide you through validation. If your photometer is out of specification for any reason, simply calibrate to your standard.

Tutorial mode can be enabled or disabled from the setup menu.

On-Screen Tutorials

The built-in tutorial mode guides users step-by-step through the measurement process. It includes all steps required for sample preparation, the required reagents, and quantities.

Tank Selection and Multiple Reading Locations Management

Users can edit up to 25 different tanks, correlate reading location to any location from the list so that logged data is linked to its reading location.

Data Logging

Missed a measurement? No need to worry because your photometer will automatically log the last 200 measurements. The results can be viewed instantly using the Log Recall Menu.


Advanced Optical System

A Light Emitting Diode (LED) provides more light while using less power and a narrow band interference filter results in more accurate and repeatable readings.

Large Cuvette Size

The 25-mm path length cuvette allows light to pass though more of the sample ensuring accurate measurements even in low-absorbance samples.

Positive Locking System

Cuvettes are placed into the holder in the same position every time by ensuring that the notch on the cap is positioned securely in the cell’s groove.

Portable Design

Made to fit comfortably in your hand for on-the-go testing or sit on a table for benchtop use, a backlit LCD display provides at-a-glance readings from any angle, day or night. The supplied AA batteries provide life for more than 800 measurements.

Ordering Information:

We believe that buying a meter should be simple. Your HI97115 photometer has many of the necessary accessories needed for successful measurements.

Each HI97115 is supplied with:

  • Sample cuvette (2 pcs.)
  • Sample cap (2 pcs.)
  • Alkaline batteries: 1.5V AA (3 pcs.)
  • Instrument quality certificate
  • A quick reference guide with instructions for manual download

HI97115C** is delivered in a rugged carrying case and is supplied with sample cuvette (2 pcs.); sample cuvette cap (2 pcs.); marine pH reagent, 30 mL dropper (1 pc.), Marine Alkalinity reagent, 30 mL bottle (1 pc.); Marine Ammonia starter kit – reagent A, 30 mL dropper (1 pc.), reagent B and C (reagent for 25 tests, each); Marine Calcium starter kit – reagent A, 30 mL bottle (1 pc.), reagent B (reagent for 25 tests); Marine Magnesium starter kit – reagent A, 120 mL bottle (1 pc.); Magnesium indicator reagent (for 25 tests); Marine Nitrate High Range reagent (reagent for 25 tests); Marine Nitrite Ultra Low Range reagent (reagent for 25 tests); Marine Phosphate Ultra Low Range reagent (reagent for 25 tests); 1 mL graduated syringe with tip (3 pcs.); minipipette with tip (1 pc.); 3 mL Pasteur pipette (2 pcs.); 5 mL syringe with black printing and tip (1 pc.); 5 mL syringe with blue printing and tip (1 pc.); 1.5V AA alkaline batteries (3 pcs.); cloth for wiping cuvettes; scissors; instrument quality certificate; and quick reference guide with instructions for manual download. * CAL Check standards sold separately. Nitrate LR testing reagent not included. ** HI97115UC is instrument code for USA.

Additional information

Weight 0.36 kg
Parameter

Chemical Oxygen Demand

Specifications

SKU HI97115-web
Name Marine Master Waterproof Wireless Multiparameter Photometer | HI97115
Marine pH Range 6.3 to 8.6 pH
Marine pH Resolution 0.1 pH
Marine pH Accuracy ±0.2 pH of reading at 25 °C
Marine pH Method Colorimetric Adaptation of the Phenol Red Method
Marine pH LED 525 nm
Marine Alkalinity Range 0.0 to 20.0 dKH
Marine Alkalinity Resolution 0.1 dKH
Marine Alkalinity Accuracy ±0.3 dKH ±5 % of reading at 25 °C
Marine Alkalinity Method Colorimetric Method
The reaction causes a distinctive range of colors from yellow to greenish blue to develop.
Marine Alkalinity LED 610 nm
Marine Ammonia Range 0.00 to 2.50 ppm (as NH₃)
Marine Ammonia Resolution 0.01 ppm
Marine Ammonia Accuracy ±0.05 ppm ±5% of reading at 25 °C
Marine Ammonia Method Adaptation of the Salicylate Method The reaction between Ammonia and Ammonium and the reagent causes a blue-green tint in the sample.
Marine Ammonia LED 610 nm
Marine Calcium Range 200 to 600 ppm
Marine Calcium Resolution 1 ppm
Marine Calcium Accuracy ±6% of reading at 25 °C
Marine Calcium Method Adaptation of the Zincon Method
Marine Calcium LED 610 nm
Marine Magnesium Range 1000 to 1800 ppm (as Mg2+)
Marine Magnesium Resolution 5 ppm
Marine Magnesium Accuracy ±5% of reading at 25 °C
Marine Magnesium Method Adaptation of the Colorimetric EDTA Method using calmagite indicator
The reaction between magnesium and the reagents causes a blue to violet tint in the sample.
Marine Magnesium LED 610 nm
Marine Nitrate LR Range 0.00 to 5.00 ppm (as NO₃–)
Marine Nitrate LR Resolution 0.01 ppm
Marine Nitrate LR Accuracy ±0.25 ppm ±2% of reading at 25 °C
Marine Nitrate LR Method Zinc Reduction Method
The reaction between nitrate and the reagent causes a pink tint in the sample.
Marine Nitrate LR LED 525 nm
Marine Nitrate HR Range 0.0 to 75.0 ppm (as NO₃–)
Marine Nitrate HR Resolution 0.1 ppm
Marine Nitrate HR Accuracy ±2.0 ppm ±5% of reading at 25 °C
Marine Nitrate HR Method Zinc Reduction Method
The reaction between nitrate and the reagent causes a pink tint in the sample.
Marine Nitrate HR LED 525 nm
Marine Nitrite ULR Range 0 to 200 ppb (as NO₂–-N)
Marine Nitrite ULR Resolution 1 ppb
Marine Nitrite ULR Accuracy ±10 ppb ±4% of reading at 25 °C
Marine Nitrite ULR Method Adaptation of the EPA Diazotization Method 354.1
The reaction between nitrite and the reagent causes a pink tint in the sample.
Marine Nitrite ULR LED 525 nm
Marine Phosphate ULR Range 0.00 to 0.90 ppm
Marine Phosphate ULR Resolution 0.01 ppm
Marine Phosphate ULR Accuracy ±0.02 ppm ±5% of reading at 25 °C
Marine Phosphate ULR Method Adaptation of Standard Methods for the Examination of Water and Wastewater, 20th Edition, Ascorbic Acid Method
Marine Phosphate ULR LED 610 nm
Measurement System Light source LED
Measurement System Bandpass filter wavelength 525 nm & 610 nm
bandwidth 8 nm wavelength accuracy ±1.0 nm
Measurement System Light detector silicon photocell
Measurement System Cuvette Type round 24.6 mm diameter (22 mm inside diameter)
Additional Specifications Auto logging 200 readings
Additional Specifications Display 128 x 64 pixel B/W LCD with backlight
Additional Specifications Auto-off after 15 minutes of inactivity (30 minutes before a READ measurement)
Additional Specifications Battery type / Life alkaline 1.5 V AA (3) / > 800 measurements (without backlight)
Additional Specifications Environment 0 to 50 °C (32 to 122 °F); 0 to 100% RH, non-serviceable
Additional Specifications Dimensions 142.5 x 102.5 x 50.5 mm (5.6 x 4.0 x 2.0 in.)
Additional Specifications Weight 380 g (13.4 oz.); with batteries
Additional Specifications Casing IP67 rating, floating

Manuals

Marine Master Waterproof Wireless Multiparameter Photometer | HI97115